I have data acquired on a Siemens scanner with head first-prone. When I convert the dicom files to MGZ, left and right and superior and inferior, respectively, are swapped. The anterior-posterior dimension is ok. I am wondering what the correct way to correct this is?

Following this entry in the wiki https://surfer.nmr.mgh.harvard.edu/fswiki/LeftRightReversal, it seems that I can use mri_convert with --in_orientation for that, and the images indeed look correct when I do this. I am however wondering whether I have to do the conversion to COR and fix the RAS coordinates as well?
I found a 2007 entry on the mailing list about that https://mail.nmr.mgh.harvard.edu/pipermail/freesurfer/2007-October/006250.html, but it does not specify how to proceed.

Your advice would be appreciated!
Thanks, Caspar