Hi Greg
I would probably say the reverse. In general mprage is optimal for CNR/unit time in terms of gray/white contrast. Flash has some advantage for e.g. distinguishing pallidum from putamen, but you would need to modify our recon stream to take advantage of them cheers Bruce
On Thu, 8 May 2014, Gregory Kirk wrote:
looking at the docs available it seems that in general flash is prefered but it seems that for cortical thickness studies you may still favor the mprage ? all with good parameter setting of course.
thanks in advance for any clarifying comments you may have
Greg _______________________________________________ Freesurfer mailing list Freesurfer@nmr.mgh.harvard.edu https://mail.nmr.mgh.harvard.edu/mailman/listinfo/freesurfer